Zone 0 / Class I Div. 1
PRESSURE SPECIALIST
TECHNICAL DATA
Ambient temperature -10 °C to +55 °C
Storage temperature -40 °C to +60 °C
Pressure input
1
/
8
in. NPT pressure tting
Pressure module input LEMO connector
Relative humidity:
(% RH operating without
condensation)
95 % (+10 to +30 °C); 75 % (+30 to +40 °C);
45 % (+40 to +50 °C); 35 % (+50 to +55 °C)
Power supply 1x 6LR61, type approved
Operating time
4 to 20 hours, depending on functions used
Dimensions (HxWxD) 216 x 94 x 66 mm (with holster)
Weight 992 g (with holster)
Specications (Summary Specications (18 °C to 28 °C, one year))
Measure pressure Model Range Resolution Accuracy Over pressure Media
Mit internem
Druckmodul
718Ex-30G -830 mbar to 2 bar 0.0001 bar 0.05 % FS 3x FS Non corrosive gases
718Ex-100G -830 mbar to 7 bar 0.001 bar 0.05 % FS 2x FS Non corrosive gases
718Ex-300G -830 mbar to 20 bar 0.001 bar 0.05 % vFS 25 bar Non corrosive gases
Using internal
pressure module
718Ex 30G
or 30G, 100G
or 300G
up to 200 bar Depending pressure module (see detailed specication)
Temp Coeff. -10 to 18 °C, 28 to 55 °C: ± 0.01 % of range per °C
Supported pressure units PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C), cm H2O (20 °C), bar, mbar, kg/cm2, mmHg, inch Hg
Source pressure Model Range
Using built in pump
718Ex 30G / 718Ex 100G / 718Ex 300G
-830 mbar to 2 bar/-830 mbar to 7 bar/-830 mbar to 20 bar
Measure current Model Range
Resolution
Accuracy
718Ex- 30G, 100G or
300G
0 - 24 mA 0.001 mA 0.02 % rdg ± 2 digits
Temp Coeff. -10 to 18 °C , 28 to 55 °C: ± 0.005 % of range per °C
ACCESSORIES
ARTICLE NO. PRODUCT DESCRIPTION
A000... Intrinsically safe pressure modules from
the 700PEx series
A0004100 DKD calibration
A0004101 WKS calibration
21